NE218
32 www.baumer.com
Change
Start count SC
➜ Press the
key.
The current start count SC is displayed for reading.
The LED „SC“ shines.
➜ Enter start count via
and keys.
Press the key.
Totalizer ∑
➜ Press the key.
The current total is displayed for reading.
The LED „∑“ shines.
If the value 999999 is exceeded, its display will be shown in two
steps. 1st step: Display of the first 6 digits. 2
nd
step: marked by an
“H“, display of 7
th
and 8
th
digit.
Display of each value for approx. 3 seconds.
Reset
➜ Press the
key.
Reset
Time meter
➜ Press the key.
The operating time is displayed for reading.
The LED „ “ shines.
➜ Press the key.
Change
Scaling factor SF
The scaling factor allows the display of fractions or multiples of the
counting pulses.
Setting range: 0.0001 to 9999.99. Setting ex factory: 1.0000.
Press the
key.
The current scaling factor „SF“ is displayed for reading.
➜ Enter scaling factor „SF“ via
and keys.
Press the
key.
Example
In the case of a length measurement by means of encoder and
cyclometer, the circumference of the cyclometer measures 200
mm and the encoder supplies 500 pulses per revolution. The mea-
surement is to be displayed in mm. The scaling factor is calculated
as follows:
Scaling factor
=
Circumference
=
200
=
0.4000
Pulses 500
Press the key again, the current count XP is re-displayed.