Operators Instruction: 11126179 00 Page 6 / 36
Operators Instruction
Electrical connection
Left side M12, 5-pin connector
1: Brown Supply (+) (15...35 vdc)
2: White Cond. (-) (4...20 mA)
3: Blue Supply (-) (15...35 vdc)
4: Black Cond. (+) (4...20 mA)
5: NC Not connected
Cond. (+) - 4
15...35 vdc (-) - 3
5 NC
1 - 15...35 vdc (+)
2 - Cond. (-)
Left side electrical connection (Front view) Right side electrical connection (Front view)
°C (+) - 4
R2 - 3
5 NC
1 - R1
2 - °C (-)
Right side M12, 5-pin connector
1: Brown R1 (external input)
2: White Temp. (-) (4...20 mA)
3: Blue R2 (external input)
4: Black Temp. (+) (4...20 mA)
5: NC Not connected
4 - Relay 2
3 - Relay 2
2 - Temp (+) (4...20 mA)
8 - R2 (external input)
Right side M12, 8-pin connector
1: White R1 (external input)
2: Brown Temp. (+) (4...20 mA)
3: Green Relay 2
4: Yellow Relay 2
5: Grey Relay 1
6: Light red Relay 1
7: Blue Temp. (-) (4...20 mA)
8: Red R2 (external input)
Right side electrical connection with relay output
Electrical connection on the display with relay output
1: Not connected
2: Not connected
3: Green Relay 2
4: Yellow Relay 2
5: Grey Relay 1
6: Light red Relay 1
(3 + 5 can be connected
common)
UnitCom Ribbon cable
to transmitter
To connect the
FlexProgrammer
COM 1 Red clip
COM 2 Black clip
DFON (UnitCom)
Supply + Temp. +
Supply - Temp. -
R1
Cond. +
R2 Cond. -
Com 1 Com 2
N.C. N.C.
Relay 2 Relay 1
Com 1 Com 2
UnitCom
Note:
If a M12 5-pin connector for left and right side is selected the
AFI4 is direct compatible with the old Baumer ISL conductivity
transmitter.
To connect the FlexProgrammer to the transmitter
Com 1 Red clip
Com 2 Black clip
The data entered to the transmitter will automatically be uploaded to
the DFON display via the ribbon cable (UnitCom)
To connect the FlexProgrammer to the DFON display
Com 1 Red clip
Com 2 Black clip
For setting the colour changes, error messages, relay set-points etc.
To set the external input for range selection
Relay 1 - 5
Relay 1 - 6
Temp. (-) (4...20 mA) - 7
R1 (external input) - 1
Range R1 R2 Range R1 R2
1 N.C. N.C. 3 N.C. 24 VDC
2 24 VDC N.C. 4 24 VDC 24 VDC
English
CombiLyz, AFI4
Note:
For ranges below 10 mS/cm a shielded cable must be used.